Customization: | Available |
---|---|
Classification: | Radiology Equipment |
Type: | CT Scanner |
Still deciding? Get samples of US$ 260000/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
* 76cm extremely large bore, more convenient for patient position and comfort.
* Tensor iteration low dose technology, accurate reconstruction, decrease misdiagnosis.
* Industry-leading dynamic real-time zoom technology, clear image details.
* 3D cone-beam parallel reconstruction technology, clearer and more realistic images.
* Dynamic reconstruction convolution technology, more powerful time resolution.
* Intelligent mA technology, lower dose.
* 80kW generator, meet all the scanning requirements.
* 80MHu Large heat capacity tube, longer lifetime, lower running cost.
* HD solid detector, fast imaging, high resolution.
* Max. 0.39s/ rotating speed when scanning, easy for joints parts examination.
* >5T superior large storage space, simple operation interface.
The image chain adopts mature technologies, the integrity of the system is balanced and strong.
The system config the industry's leading dynamic real-time zooming technology and real 3D cone parallel reconstruction technology, so that the system can get the aperture expanded and at the same time get high definition image and leading high resolution image quality.
Specifications | ||
No. | Item | Specification |
1 | Gantry | |
1.1 | Patient Opening | 76cm |
1.2 | Gantry Driven | Motor-driven |
1.3 | Tilt Capability | ±30° |
1.4 | Gantry Remote-Control | Yes |
1.5 | Detector Type | GOS high-definition solid detector |
1.6 | Distance of Focus-Isocenter | 570mm |
1.7 | Numbers of Detector Rows | 32 |
1.8 | Width of Z-Axle Detector | 20mm |
1.9 | Detector Columns of Channels per Row | 864 |
1.1 | Numbers of Detector Columns | 27648 |
1.11 | Slice Number | 64 |
1.12 | DAS Column Number | 55296 |
1.13 | DAS Acquisition Modes | 64×0.625mm |
32×0.625mm | ||
16×1.25mm | ||
8×2.5mm | ||
4×5.0mm | ||
2×10mm | ||
1.14 | Data-Transfer Type | RF, optical fiber communication |
1.15 | 3D Laser Orientation | Provided |
1.16 | Automatic Exposure Control | Provided |
1.17 | Auto-Voice Function | Yes |
2 | Scan Parameter | |
2.1 | Shortest 360 Degree Rotation Time | 0.39s |
2.2 | Allowed Rotation Times | 0.39s, 0.5s,0.6s, 0.7s, 0.8s,1.0s,2.0s |
2.3 | Slice Numbers of Rotation | 64 |
2.4 | Minimum Scan Slice Thickness | 0.625mm |
2.5 | Minimum Reconstruction Thickness | 0.625mm |
2.6 | Maximum Scan Slice Thickness | 10mm |
2.7 | Reconstruction Slice Thickness | 0.625mm |
2.8 | Speed of Image Reconstruction | 10 frames/s |
2.9 | Scan FOV | 50cm |
2.1 | Max. Image Reconstruction Matrix | 1024×1024 |
2.11 | Max. Image Display Matrix | 1024×1024 |
2.12 | Maximum Continuous Scan Duration | 100s |
2.13 | Maximum Continuous Scan Length | 1950mm |
2.14 | Range of Pitch | 0.13-1.5 |
2.15 | Scan Mode | Scout Scan |
Axial Scan | ||
Helical Scan | ||
Cine Scan | ||
3 | HV generator and Tube | |
3.1 | Maximum Output Power of Generator | 80kW |
3.2 | Tube kV Selections | 80kV,100kV |
120kV,140kV | ||
3.3 | Tube mA Range | 10~667mA |
3.4 | Tube Heat Capacity | 8.0MHu |
3.5 | Heat Dissipation Rate | 931 kHU/min |
3.6 | Type of Cooling | Oil cooling + Air cooling |
3.7 | Tube Focus | 1.1mm×1.2mm |
0.6mm×1.2mm | ||
3.8 | Flying Focal Spot Technology | Provided |
3.9 | mA Modulation Technology | Provided |
4 | Patient Table | |
4.1 | Can Position Outside of Scan Room | Yes |
4.2 | Laser Position System Accuracy | ±1mm |
4.3 | Maximum Horizontal Movable Range | 1950mm |
4.4 | Table Horizontal Scan Range | 1850mm |
4.5 | Table accuracy | 0.25mm |
4.6 | Table Vertical Movable Range | 425~990mm |
4.7 | Maximum Speed of Vertical Movement | 16.5mm/s |
4.8 | Maximum Speed of Horizontal Movement | 5 mm/s ~ 160 |
4.9 | Maximum Patient Weight | 250kg |
4.1 | Foot Pedal of Table | Optional |
5 | Image Quality | |
5.1 | High Contrast Resolution | ≥20lp/cm (0% MTF) |
5.2 | Low Contrast Resolution | [email protected]% (31mGy) |
5.3 | Isotropic Imaging Resolution | 0.3mm |
5.4 | Range of CT Value | -1000~+1000 |
5.5 | Image Noise | ≤70dB |
6 | Computer | |
6.1 | CPU | 4coreX3.5Hz |
6.2 | Memory | 32GB |
6.3 | Storage of Hard-Disk | 5TB |
6.4 | Monitor | 24'' LCD Monitor |
6.5 | Resolution of Monitor | 1920 x 1200 |
6.6 | Image-Data External Storage Type | CD/DVD/USB |
6.7 | DICOM 3.0 | Provided |
6.8 | Work list | Provided |
6.9 | 3D Post-Processing | MPR, CPR, SSD, MIP, VR, CTA |
6.1 | Contrast Agent Automatically Tracking and Trigger Technology | Provided |
6.11 | X-ray intelligent control technology | Provided |
6.12 | Maximum CTA Scan Range | 1850mm |
6.13 | Beam Artifacts Correction Technology | Provided |
6.14 | Posterior Fossa Image Optimization Technology | Provided |
6.15 | Metal Artifacts Algorithm | Provided |
6.16 | Scalpel Simulate Technology | Provided |
6.17 | Child Mode | Provided |
6.18 | One Key VR Technology | Provided |
6.19 | CTVE Technology | Provided |