Customization: | Available |
---|---|
Classification: | Imaging Diagnostic Equipment |
Type: | X Ray Equipment |
Still deciding? Get samples of US$ 3850/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Item | Content | Index |
Demand of power supply | Voltage | Single phase 220V/380V+/-10% |
Frequency | 50HZ+/-1HZ | |
Capacity | 10kVA | |
Internal Resistance | 220V:0.6Ω 380V:2Ω | |
Photography | Voltage | 50-90kV |
Current | Small focus: 16mA | |
Big focus: 32mA, 63mA, 100mA | ||
Time | 0.04-6.3s, 22 grades together | |
High Voltage transformer unit | Capacity | 7kVA |
Max. DC output voltage | 90kV | |
Max. DC output current | 100mA | |
X-Ray tube | Model | XD4-2.9/100 |
Focus | Small focus: 1.8mm | |
Big focus: 4.3mm | ||
Photographic bed | Bed Face (L*W*H) | 1900mmx695mmx700mm |
Ray-filter | Grid density: N28 | |
Convergence distance: 100cm | ||
Grid ratio: r8 | ||
Pillar of X-Ray source unit | Moving in length | >1350mm |
Moving in breadth | >360mm | |
Moving up and down | 720-1800mm | |
Max size cassette for photograph | 365mmx432mm (14"x17") | |
Gross weight | 700kg | |
Net weight | 550kg | |
Dimensions | 2180mmx800mmx1130mm |