Feature:
* Air temp.and skin temp. servo-controlled by computer
* Various and self-check alarms
* Free-step mattress tilting adjustment
* Removable humidity reservoir, easy to clean
* >37ºC temperature setting
* With the transfusion shelf and tray
* Added with guardrail
* RS-232 connector
* Phototherapy unit (option)
| Data |
|
| Electrical requirement |
~220V 50Hz or ~ 230V 60Hz |
| Power consumption |
≤ 450VA |
| Air mode temp. range |
25.0ºC~ 37.0ºC |
| Skin mode temp. range |
34.0ºC~ 37.0ºC |
| Air over temp. range |
< 38.0ºC |
| Air mode high temp. range |
37.1ºC~ 39.0ºC |
| Skin mode high temp. range |
37.1ºC~ 38.0ºC |
| High temp. mode alarm |
< 40.0ºC |
| Air mode alarm |
±3.0ºC |
| Skin mode alarm |
±1.0ºC |
| Skin sensor accuracy |
≤ ±0.3ºC |
| Air control accuracy |
≤ 0.5ºC |
| Mattress temp. uniformity |
≤ 0.8ºC |
| Noise |
≤ 55dB(A) |