Customization: | Available |
---|---|
Classification: | Medical Laboratory Equipment |
Type: | Biological Safety Cabinet |
Still deciding? Get samples of US$ 3500/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Features:
1. Color key display, dynamicly display various values of the safety cabinet, and has appointment time function. Automatically adjust the wind speed.
2. With power-off memory function. the state before power-off will be restored after power-on, and there will be visual prompts and alarms.
3. The cabinet is separated from the base. The height of the adjustable base can be customized.
4. Interlocking function: UV lamp and front window; UV lamp and fan, lamp; fan and front window.
5. The front window closes the double trigger signal, so that the sterilization and disinfection function of the ultraviolet lamp is normally turned on.
6. The negative pressure air duct is equipped with a foreign matter filtering structure to prevent foreign matter such as paper scraps from entering the fan system and affecting the normal operation.
7. The front window glass has one-key lit function.
8. Designed with a 10° inclination angle, which is in line with ergonomic principles, with a larger viewing angle.
Technical Data | |
Model | LSC1100IIA2-X |
External Size(W*D*H) | 1100*755*2200mm |
Internal Size(W*D*H) | 940*600*660mm |
Tested Opening | Safety Height 200mm(8") |
Max Opening | 440mm |
Inflow Velocity | 0.53±0.025m/s |
Downflow Velocity | 0.33±0.025m/s |
ULPA Filter | Two, 99.9995% efficiency at O.12μm, filter life indicator |
Front Window | Motorized. Two-layer laminated toughened glass≥6mm. Anti UV |
UV Lamp | 30W*1 UV timer, UV life indicator, emission of 235.7 nanometers for most efficient decontamination |
LED Lamp | 12W*2 |
Illumination | ≥1000Lux |
Power | 600W |
Waterproof Socket | Two, total load of two sockets: 500W |
Display | LCD display: exhaust filter and downflow flter pressure, filter and UV lamp working time, inflow and downflow velocity, filter life, humidity and temperature,system working time etc. |
Airflow System | 70% air recirculation. 30% air exhaust |
Visual and Audio Alarm | Abnormal airlow velocity, filter replacement, front window at unsafe height, high fiter pressure alarm, abnormal power failure |
Work Surface Height | 750mm (Size can be customized according to requirements) |
Caster | Footmaster caster |
Power Supply | AC220V 50160Hz; 110V 60Hz |
Standard Accessory | LED lamp 2pcs, UV lamp 1pc, Base stand, Remote control, Foot switch, Drain valve, Waterproof socket 2pcs |
Optional Accessory | Water and gas tap, Electric height adjustable base stand, Armrest |
Gross Weight | 220kg |
Package Size(W*D*H) | 1230*1060*1840mm |